How to solve these minor problems with mercury analyzers?

The mercury analyzer is a highly sensitive instrument for measuring mercury content. It uses the cold atomic absorption method to measure the mercury content in liquids, solid samples converted into liquids after treatment, or mercury content in gas samples.

 mercury content detector

How to troubleshoot these minor problems with lab mercury analyzer?

1. No display after powering on: Replace the fuse.

2. Mercury lamp indicator light not on: Restart the analyzer.

3. Low sensitivity:

① Incorrect flow rate or expired mercury standard sample.

② Air leak in the pipeline.

③ Turn the display adjustment counterclockwise all the way to the end.

4. High sensitivity:

① Incorrect flow rate.

② Pipeline contamination.

③ Turn the display adjustment clockwise all the way to the end.

5. Poor reproducibility: Improper operation, pipeline contamination, desiccant problem, pipeline leak.

 mercury gas analyzer

Instruments as the Future-Shaping Core Engine Advance Scientific Exploration to New Frontiers

The precision of high-end instruments means an infinite approach to scientific truth. NANBEI INSTRUMENT strictly controls every step of the manufacturing process, achieving a production pass rate over 99.5%.


Online customer service

  • Mobile: +86 15890068607
  • Tel: +86-371-60310701
  • Skype: angelina72811
  • WhatsApp: +86 15890668062
  • Email: bella@nanbei-china.com
  • Add: No. 72, Beihuan Road, Jinshui District, Zhengzhou,China

Feedback Online

Msg Title:

*E-mail:

Message: