The mercury analyzer is a highly sensitive instrument for measuring mercury content. It uses the cold atomic absorption method to measure the mercury content in liquids, solid samples converted into liquids after treatment, or mercury content in gas samples.

How to troubleshoot these minor problems with lab mercury analyzer?
1. No display after powering on: Replace the fuse.
2. Mercury lamp indicator light not on: Restart the analyzer.
3. Low sensitivity:
① Incorrect flow rate or expired mercury standard sample.
② Air leak in the pipeline.
③ Turn the display adjustment counterclockwise all the way to the end.
4. High sensitivity:
① Incorrect flow rate.
② Pipeline contamination.
③ Turn the display adjustment clockwise all the way to the end.
5. Poor reproducibility: Improper operation, pipeline contamination, desiccant problem, pipeline leak.